Apologies and Correction for the product specification description of Data Logger KEW 5010/5020
Thank you for using our products and services.
Regarding the normal recording mode of the Data Logger KEW 5010/5020, the instruction manual and other documents stated that "record the average measurement value at each recording interval", but it is not an average value but is designed to “record the measured values at every recording interval”. Therefore, we modify the descriptions in the instruction manual to match the actual specifications.
- Affected Products
Data Logger KEW 5010/5020 - Details
In normal recording mode, sampling is performed at approximately 1.6 ms intervals. True RMS values are calculated approximately every 100 ms, and the average of 10 True RMS values obtained during each 1 sec period is used as the measurement value.
In the instruction manual and other documents stated that “average of measurement values is recorded at each recording interval”, however, “the measured values are recorded at every recording interval” with the actual specification. Therefore, if the number of measurement values increases or decreases between the preset recording intervals, slight differences may occur.
Wrong: Record the average of measurement values
(explanation in manuals, etc.)
Correct: Record measurement values at each recording interval
(actual specification) - Defective Product Handling
In accordance with the actual specifications, all explanations related to the normal recording mode will be revised and updated on our website, in our catalog, and in the instruction manual. The updated instruction manual can be viewed by clicking here.
The updated instruction manual can be viewed by clicking here.
We apologize in advance for any inconvenience this may cause and appreciate your understanding.
KYORITSU ELECTRICAL INSTRUMENTS WORKS, LTD.
Quality Assurance Department